Suscribirse

Inhalation device requirements for patients' inhalation maneuvers - 29/08/16

Doi : 10.1016/j.rmed.2016.07.013 
Peter Haidl a, Stefan Heindl b, Karsten Siemon a, Maria Bernacka c, Rolf Michael Cloes c,
a Hospital Kloster Grafschaft, Germany 
b Bahnhofstraße 22, 82131 Gauting, Germany 
c Mundipharma GmbH, Germany 

Corresponding author. Mundipharma GmbH, Mundipharmastraße 2, 65549, Limburg, Germany.Mundipharma GmbHMundipharmastraße 2Limburg65549Germany

Abstract

Background

Inhaled drugs are the mainstay of treatment for lung diseases such as asthma and chronic obstructive pulmonary disease. However, failure to use inhalation devices correctly can lead to a poorly controlled status. A vast number of inhalation devices exist and each device has specific requirements to achieve optimum inhalation of the drug. Currently, there is no overview of inhalation requirements considering all devices. This article presents a review of the literature on different inhalation device requirements and incorporates the data into a new inhalation flow algorithm.

Methods

Data from literature on commercially available inhalation devices were evaluated and parameters, such as inhalation flow rate, flow acceleration, inhalation volume, and inspiration time assessed for the required inhalation maneuver specific to the device. All agreed upon data points were used to develop an inhalation flow algorithm.

Results

The literature analysis revealed availability of robust data for the required inhalation flow characteristics for most devices and thus for the development of an algorithm. For those devices for which these parameters are not published, the minimum required flow criteria were defined based on published data regarding individual aspects of aerosol quality.

Conclusions

This review provides an overview of inhalation devices available on the market regarding requirements for an acceptable inhalation maneuver and shows which goals should be achieved in terms of inhalation flows. The presented algorithm can be used to develop a new computer based measurement system which could help to test and train patients' individual inhalation maneuvers with their inhalation devices.

El texto completo de este artículo está disponible en PDF.

Highlights

Limit values of inspiratory flow requirements for inhalation devices are presented.
Information on flow acceleration, inspiratory time, inspiratory volume is given.
New measuring algorithm enables evaluation of individual inhalation maneuvers.

El texto completo de este artículo está disponible en PDF.

Keywords : Inhalation device, Inspiratory flow rate, Inhalation maneuver, Inhalation maneuver quality, Drug delivery


Esquema


© 2016  The Authors. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 118

P. 65-75 - septembre 2016 Regresar al número
Artículo siguiente Artículo siguiente
  • Asthma prevalence in German Olympic athletes: A comparison of winter and summer sport disciplines
  • Charlotte Selge, Silke Thomas, Dennis Nowak, Katja Radon, Bernd Wolfarth

¿Ya suscrito a @@106933@@ revista ?